Varian / Bruker Series 450-GC EFC 21 SI (740992-M)

Located on the right rear of the GC. Electronic control of the inlet pneumatics.

Category: Tags: , , , , ,

Troubleshooting


Possible issues caused by this broken part:

  • Varian / Bruker Series 450-GC Activates Safety Method.
  • Varian / Bruker Series 450-GC Pneumatics problems.

Possible errors prompted by this broken part:

Error codes:

The following is a list of errors codes concerning the EFC:

 

Error Code & Description Possible causes Suggested actions
35 – 37 Front-middle-back Flow Out of Range Flow is above max+10% or below – 5kPa for more then 10 seconds. Check EFC flows (gas supply).
38 – 40 Front-middle-back Pressure Out of Range Pressure is above max+10% or below -5kPa for more then 10 seconds.
41 – 43 Front-middle-back EFC Flow Ready Time Out Unit not ready for 10 minutes. Check gas supply, set desired pressure, check EFC device, reboot GC.
44 – 46 Front-middle-back EFC Pressure Ready Time Out Unit not ready for 10 minutes.
304 – 306 Front EFC Channel 1-2-3 Out of Range Flow is above max +10 % or below – 5 mL/min for more then 10 sec. Turn off/on flow, check source (gas-supply)
307 – 309 Middle EFC Channel 1-2-3 Out of Range Flow is above max +10 % or below – 5 mL/min for more then 10 sec. Turn off/on flow, check source (gas-supply)
310 – 312 Back EFC Channel 1-2-3 Out of Range Flow is above max +10 % or below – 5 mL/min for more then 10 sec. Turn off/on flow, check source (gas-supply)
313 – 315 Front EFC Chanel 1-2-3, Flow Ready Time Out Unit not ready for 10 minutes
316 – 318 Middle EFC Chanel 1-2-3, Flow Ready Time Out Unit not ready for 10 minutes
319 – 321 Back EFC Chanel 1-2-3, Flow Ready Time Out Unit not ready for 10 minutes