Troubleshooting
Possible issues caused by this broken part:
- Varian / Bruker Series 450-GC doesn’t work / turn on.
Possible errors prompted by this broken part:
Error codes:
The following is a list of errors codes concerning the Main Board. Any of these indicate a defective Main Board and should be changed:
- 444 – P10V Ref Out of Tolerance
- 445 – AIN10V Out of Tolerance
- 446 – P5V Ref Out of Tolerance
- 447 – P2V5 Ref Out of Tolerance
- 448 – P24V Prim Out of Tolerance
- 449 – P24V Sec Out of Tolerance
- 450 – N24V Out of Tolerance
- 451 – P5V D Out of Tolerance
- 452 – P5V Out of Tolerance
- 453 – N5V Out of Tolerance
- 454 – P15V Out of Tolerance
- 455 – N15V Out of Tolerance
- 456 – P5V DAC Out of Tolerance
- 457 – VREF1 Out of Tolerance
- 458 – VREF0 Out of Tolerance