| Message | 
Probable causes | 
Suggested actions | 
| 2014 / 2500 – Pressure above upper limit | 
- Upper pressure limit set too low.
 
- Blockage in the flowpath (after the damper).
 
- Defective damper.
 
- Defective main board.
 
 
 | 
- Ensure the upper pressure limit is set to a value suitable for the analysis.
 
- Check for blockage in the flowpath. The following components are particularly subject to blockage: inline filter frit, needle (autosampler), seat capillary (autosampler), sample loop (autosampler), column frits and capillaries with small internal diameters (e.g. 50 μm ID).
 
- Contact Moyo Scientific Service
 
 
 | 
| 2015 / 2501 – Pressure below lower limit | 
- Lower pressure limit set too high.
 
- Air bubbles in the mobile phase.
 
- Leak.
 
- Defective damper.
 
- Defective main board.
 
 
 | 
- Ensure the lower pressure limit is set to a value suitable for the analysis.
 
- Make sure that the degasser is in flow path and works correctly. Purge the module.
 
- Ensure solvent inlet filters are not blocked.
 
- Inspect the pump head, capillaries and fittings for signs of a leak.
 
- Purge the module. Run a pressure test to determine whether the seals or other module components are defective.
 
- Contact Moyo Scientific Service.
 
 
 | 
| 2044 – AIV fuse | 
- Defective active inlet valve.
 
- Defective connection cable (front panel to main board).
 
- Defective main board.
 
 
 | 
- Restart the module. If the error message appears again, exchange the active inlet valve.
 
- Contact Moyo Scientific Service.
 
 
 | 
| 2041 / 2042 – Motor-drive power | 
- Flow path blockage in front of the damper.
 
- Blocked passive inlet valve.
 
- Blocked outlet valve.
 
- High friction (partial mechanical blockage) in the pump drive assembly.
 
- Defective pump drive assembly.
 
- Defective main board.
 
 
 | 
- Ensure the capillaries and frits between the pump head and damper inlet are free from blockage.
 
- Exchange the passive inlet valve.
 
- Exchange the outlet valve.
 
- Remove the pump-head assembly. Ensure there is no mechanical blockage of the pump-head assembly or pump drive assembly.
 
- Contact Moyo Scientific Service
 
 
 | 
| 2517 – Temperature limit Exceeded | 
- High friction (partial mechanical blockage) in the pump drive assembly.
 
- Partial blockage of the flowpath in front of the damper.
 
- Defective pump drive assembly.
 
- Defective main board.
 
 
 | 
- Remove the pump-head assembly. Ensure there is no mechanical blockage of the pump-head assembly or pump drive assembly.
 
- Ensure the outlet valve is not blocked.
 
- Contact Moyo Scientific Service.
 
 
 | 
| 2201 / 2211 – Servo restart failed | 
- Disconnected or defective cable.
 
- Blocked passive inlet valve.
 
- Mechanical blockage of the module.
 
- Defective pump drive assembly.
 
- Defective main board.
 
 
 | 
- Exchange the passive inlet valve.
 
- Remove the pump-head assembly. Ensure there is no mechanical blockage of the pump-head assembly or pump drive assembly.
 
- Contact Moyo Scientific Service.
 
 
 | 
| 2202 / 2212 – Pump head missing | 
- Pump head not installed correctly (screws not secured, or pump head not seated correctly).
 
- Broken piston.
 
 
 | 
- Install the pump head correctly. Ensure nothing (e.g. capillary) is trapped between the pump head and body.
 
- Exchange the piston.
 
 
 | 
| 2203 / 2213 – Index limit | 
- Irregular or sticking drive movement.
 
- Defective pump drive assembly.
 
 
 | 
- Remove the pump head, and examine the seals, pistons, and internal components for signs of wear, contamination or damage. Exchange components as required.
 
- Contact Moyo Scientific Service
 
 
 | 
| 2204 / 2214 – Index adjustment | 
- Irregular or sticking drive movement.
 
- Defective pump drive assembly.
 
 
 | 
- Remove the pump head, and examine the seals, pistons, and internal components for signs of wear, contamination or damage. Exchange components as required.
 
- Contact Moyo Scientific Service.
 
 
 | 
| 2205 / 2215 / 2505 – Index missing | 
- Disconnected or defective encoder cable.
 
- Defective pump drive assembly.
 
 
 | 
Contact Moyo Scientific Service. | 
| 2206 / 2216 – Stroke length | 
Defective pump drive assembly. | 
Contact Moyo Scientific Service. | 
| 2207 / 2217 – Initialisation failed | 
- Blocked passive inlet valve.
 
- Defective pump drive assembly.
 
- Defective main board.
 
 
 | 
- Exchange the passive inlet valve.
 
- Contact Moyo Scientific Service.
 
 
 | 
| 2053 – Wait timeout | 
- Purge valve open.
 
- Leak at fittings, purge valve, active inlet valve, outlet valve or piston seals.
 
- Flow changed after starting test.
 
- Defective pump drive assembly.
 
 
 | 
- Ensure that purge valve is closed.
 
- Ensure pump components are seated correctly. If there are still signs of a leak, exchange the appropriate seal (purge valve, active inlet valve, outlet valve, piston seal).
 
- Exchange defective capillaries.
 
- Ensure correct operating condition for the special application in use.
 
- Contact Moyo Scientific Service.
 
 
 |