| Message | 
Probable cause | 
Suggested actions | 
| Bottle in gripper | 
- The sample vial was not delivered properly and stayed in the tray gripper.
 
 
 | 
- Remove the vial and return it to its position in the tray.
 
- Ensure that the tray quadrants are snapped into place.
 
- Ensure that the injector is plugged into the correct connector on the back of the GC. For 6890, ensure it is configured properly.
 
- Check the deliver to location for the vial and verify that the location is empty and free from obstructions.
 
- Make sure that the injector is upright/vertical on the GC.
 
- Restart the sequence.
 
- If the error occurs again:
 
Replace Autosampler Vial Gripper 
Replace Autosampler Spring Loaded Belt Tensor 
OR 
Contact Moyo Scientific Service. 
 
 | 
| Front (or Back) door open or Injector not mounted | 
 | 
- See “Fault light is on” in Faults
 
 
 | 
| Front (or Back) injector com error | 
- There is a communications error between the injector and the GC.
 
 
 | 
 | 
| Front (or Back) Injector Incomplete Injection | 
- The syringe needle is bent.
 
- The plunger or syringe carriage is operating incorrectly during injection.
 
 
 | 
- See “Correcting Syringe Problems” on page 24 of the user manual.
 
- Remove the syringe from the injector and check the plunger for stickiness or binding. Replace the syringe if necessary.
 
- Restart the sequence.
 
- If the error occurs again,
 
Replace Injector Needle Assembly. 
Replace Injector Needle Support Rod. 
OR 
Contact Moyo Scientific Service. 
 
 | 
| Front (Or Back) Injector Reset | 
- There is an interruption in the power supply from the GC
 
 
 | 
- Contact Moyo Scientific Support
 
 
 | 
| Front (Or Back) Plunger Error | 
- The syringe plunger is sticking, or not securely connected to the plunger carrier.
 
- The plunger solenoid is binding.
 
- The plunger carrier encoder is inoperable.
 
 
 | 
- Remove the syringe and check it for plunger stickiness or binding. Replace the syringe if necessary. For more information, see Inspecting a syringe in the Operation Guide.
 
- Check the viscosity of the sample against the viscosity parameter. Reset the viscosity parameter if necessary.
 
- Restart the sequence.
 
- If the error occurs again,
 
Replace Injector Plunger Motor. 
Replace Injector Needle Assembly. 
OR 
Contact Moyo Scientific Service. 
 
 | 
| Front (Or Back) Syringe Error | 
- The syringe carriage motor is defective.
 
- The syringe is not currently installed or is an incorrect type.
 
- The syringe carriage sensor is inoperable.
 
 
 | 
- Ensure the syringe is installed correctly. For more information, see your Operation Guide.
 
- Ensure the syringe meets specifications.
 
- If the syringe needle is bent, see Correcting Syringe Problems on page 24 of the user manual.
 
- Restart the sequence.
 
- If the error occurs again, contact Moyo Scientific Service
 
 
 | 
| Front (Or Back) Turret Error | 
- Something has interfered with the turret rotation.
 
- The turret motor/encoder assembly is inoperable.
 
- The turret type was changed while the power was on and the turret alignment procedure was not performed.
 
- Turret is loose.
 
 
 | 
- Clear any obstructions.
 
- Check the Align Mode light. If it is lit, perform the alignment procedure (See Turret alignment in the user manual).
 
- Tighten the knurled nut on the top of the turret.
 
- If the error occurs again,
 
Replace Injector Turret. 
Replace Injector Turret Motor. 
OR 
Contact Moyo Scientific Service. 
 
 | 
| Injector Not Present | 
- There is a board failure in the injector or GC.
 
- The injector cable is bad or not securely connected to the GC.
 
- There is a cable failure in the GC.
 
- Your method specifies an incorrect injector location (method mismatch).
 
 
 | 
- Make sure that the injector to GC cable connection is secure.
 
- Check your method to make sure it uses the appropriate injector location.
 
- If the error remains,
 
Replace GC Mainboard. 
Replace Injector Rear Board (G2613-60010-M). 
OR 
Contact Moyo Scientific Service. 
 
 | 
| Injector Offline | 
- There is a board failure in the injector or GC.
 
- The injector cable is bad or not connected.
 
- There is a cable failure in the GC.
 
 
 | 
- Make sure that the injector to GC cable connection is secure.
 
- If the error remains,
 
Replace GC Mainboard. 
Replace GC Injector Rear Board. 
OR 
Contact Moyo Scientific Service. 
 
 | 
| No Bar Code Reader | 
- Bar code reader cable not securely connected.
 
- Bar code reader is defective.
 
- Tray is defective.
 
 
 | 
- Ensure the bar code reader cable connection is secure.
 
- If the problem continues, Contact Moyo Scientific Service.
 
 
 | 
| No Bottle in Gripper | 
- The sample vial was not found by the gripper.
 
- The gripper could not grasp the vial.
 
- The vial was dropped during transfer to or from the turret.
 
- The sensor in the gripper is defective
 
 
 | 
- Make sure that the sample vials are in the locations specified by the sequence.
 
- Make sure that the sample vials meet recommended specification. See your Sampling Techniques Handbook.
 
- If you are using adhesive labels, verify that the labels are properly installed. See your Operation Guide.
 
- If the error recurs frequently,
 
Replace Autosampler Vial Gripper. 
Replace Autosampler Arm Board. 
Replace Autosampler Flex Cable. 
OR 
Contact Moyo Scientific Service. 
 
 | 
| Tray Not Present | 
- There is a board failure in the tray or GC.
 
- The tray cable is bad or not connected between the GC and tray.
 
- There is a cable failure in the GC.
 
 
 | 
- Make sure that the tray cable connection is secure.
 
- Replace the tray cable.
 
- If the error remains:
 
Replace GC Mainboard. 
OR 
Contact Moyo Scientific Service. 
 
 | 
| Tray Offline | 
- There is a board failure in the tray or GC.
 
- The tray cable is bad or not connected.
 
- There is a cable failure in the GC
 
 
 | 
- Make sure that the tray to GC cable connection is secure.
 
- Replace the tray cable.
 
- If the error remains:
 
Replace GC Mainboard. 
OR 
Contact Moyo Scientific Service. 
 
 | 
| Invalid Sequence | 
- The sequence is set up for the wrong injection device.
 
- Hardware required by the sequence is not installed and configured.
 
- The GC configuration was changed during sequence execution.
 
- The injector cable is bad or not connected properly
 
 
 | 
- Make sure that the connection to the GC is secure.
 
- Verify the sequence parameters against the GC configuration.
 
- If the error remains, contact Moyo Scientific Service.
 
 
 | 
| No Injector | 
- The cabling connection to the GC became loose during a run.
 
- An injector board or GC board failed during a run.
 
 
 | 
- Make sure that the connection to the GC is secure.
 
- If the error remains:
 
Replace GC Mainboard. 
Replace Injector Read Board. 
OR 
Contact Moyo Scientific Service 
 
 | 
| Prerun > 10 Min | 
 | 
- Check for Not Ready and other GC messages to determine the cause
 
 
 | 
| Sampler Error | 
- The sampler could not function for an undocumented reason
 
 
 | 
- Record the code number shown in the Control Module error message.
 
If the problem persists, Contact Moyo Scientific Service and report the error number 
 
 |