Agilent Series 5973 / 5975 MSD High Vacuum Pump Fan (3160-1037-M)

Fan located at the front of the G2589A MSD, pointing towards the High Vacuum Pump.

Troubleshooting:

Possible issues caused by this broken part:

  • Agilent 5973 inert MSD shuts down
  • Agilent 5973 inert MSD doesn’t reach vacuum
  • Agilent 5973 inert MSD can’t keep vacuum
  • Agilent 5973 inert MSD High Vacuum Pump turns off

Possible errors prompted by this broken part:

VACUUM SYMPTOMS:
  • Pumpdown failure shutdown
ERROR MESSAGES:
  • Difficulty with the high vacuum pump
  • Foreline pressure has exceeded 300 mTorr
  • The high vacuum pump is not ready
  • The system is in vent state