Troubleshooting
Possible issues caused by this broken part:
- Agilent 5973 MSD doesn’t work.
- Agilent 5973 MSD shuts down.
- Agilent 5973 MSD poor quality
- Agilent 5973 MSD poor / strange results
Possible errors prompted by this broken part:
CHROMATOGRAPHICS SYMPTOMS:
- Baseline wanders
- Does not pass tune
- Poor sensitivity
- Poor repeatability
MASS SPECTRAL SYMPTOMS:
- No peaks
- Isotopes are missing or isotope ratios are incorrect
- High background
- Peaks have precursors
- Peaks widths are inconsistent
- Relative abundance of m/z 502 is less than 3%
- High mass sensitivity is poor
ERROR MESSAGES:
- Lens supply fault
- No peaks found
- Temperature control disabled
- Temperature control fault
- There is no emission current
- There is not enough signal to begin tune